A line fiducial method for geometric calibration of cone-beam CT systems with diverse scan trajectories

M. W. Jacobson, M. D. Ketcha, S. Capostagno, A. Martin, A. Uneri, J. Goerres, T. De Silva, S. Reaungamornrat, R. Han, A. Manbachi, J. W. Stayman, S. Vogt, G. Kleinszig, J. H. Siewerdsen

Research output: Contribution to journalArticle

Abstract

Modern cone-beam CT systems, especially C-arms, are capable of diverse source-detector orbits. However, geometric calibration of these systems using conventional configurations of spherical fiducials (BBs) may be challenged for novel source-detector orbits and system geometries. In part, this is because the BB configurations are designed with careful forethought regarding the intended orbit so that BB marker projections do not overlap in projection views. Examples include helical arrangements of BBs (Rougee et al 1993 Proc. SPIE 1897 161-9) such that markers do not overlap in projections acquired from a circular orbit and circular arrangements of BBs (Cho et al 2005 Med. Phys. 32 968-83). As a more general alternative, this work proposes a calibration method based on an array of line-shaped, radio-opaque wire segments. With this method, geometric parameter estimation is accomplished by relating the 3D line equations representing the wires to the 2D line equations of their projections. The use of line fiducials simplifies many challenges with fiducial recognition and extraction in an orbit-independent manner. For example, their projections can overlap only mildly, for any gantry pose, as long as the wires are mutually non-coplanar in 3D. The method was tested in application to circular and non-circular trajectories in simulation and in real orbits executed using a mobile C-arm prototype for cone-beam CT. Results indicated high calibration accuracy, as measured by forward and backprojection/triangulation error metrics. Triangulation errors on the order of microns and backprojected ray deviations uniformly less than 0.2 mm were observed in both real and simulated orbits. Mean forward projection errors less than 0.1 mm were observed in a comprehensive sweep of different C-arm gantry angulations. Finally, successful integration of the method into a CT imaging chain was demonstrated in head phantom scans.

Original languageEnglish (US)
Article number025030
JournalPhysics in Medicine and Biology
Volume63
Issue number2
DOIs
StatePublished - Jan 1 2018

Fingerprint

Cone-Beam Computed Tomography
Orbit
Calibration
Arm
Head

Keywords

  • BB phantom
  • C-arm
  • cone-beam CT
  • geometric calibration
  • line fiducial
  • wire phantom

ASJC Scopus subject areas

  • Radiological and Ultrasound Technology
  • Radiology Nuclear Medicine and imaging

Cite this

A line fiducial method for geometric calibration of cone-beam CT systems with diverse scan trajectories. / Jacobson, M. W.; Ketcha, M. D.; Capostagno, S.; Martin, A.; Uneri, A.; Goerres, J.; De Silva, T.; Reaungamornrat, S.; Han, R.; Manbachi, A.; Stayman, J. W.; Vogt, S.; Kleinszig, G.; Siewerdsen, J. H.

In: Physics in Medicine and Biology, Vol. 63, No. 2, 025030, 01.01.2018.

Research output: Contribution to journalArticle

Jacobson, MW, Ketcha, MD, Capostagno, S, Martin, A, Uneri, A, Goerres, J, De Silva, T, Reaungamornrat, S, Han, R, Manbachi, A, Stayman, JW, Vogt, S, Kleinszig, G & Siewerdsen, JH 2018, 'A line fiducial method for geometric calibration of cone-beam CT systems with diverse scan trajectories', Physics in Medicine and Biology, vol. 63, no. 2, 025030. https://doi.org/10.1088/1361-6560/aa9910
Jacobson, M. W. ; Ketcha, M. D. ; Capostagno, S. ; Martin, A. ; Uneri, A. ; Goerres, J. ; De Silva, T. ; Reaungamornrat, S. ; Han, R. ; Manbachi, A. ; Stayman, J. W. ; Vogt, S. ; Kleinszig, G. ; Siewerdsen, J. H. / A line fiducial method for geometric calibration of cone-beam CT systems with diverse scan trajectories. In: Physics in Medicine and Biology. 2018 ; Vol. 63, No. 2.
@article{9797dd2b87214afab2677d5d7b782c85,
title = "A line fiducial method for geometric calibration of cone-beam CT systems with diverse scan trajectories",
abstract = "Modern cone-beam CT systems, especially C-arms, are capable of diverse source-detector orbits. However, geometric calibration of these systems using conventional configurations of spherical fiducials (BBs) may be challenged for novel source-detector orbits and system geometries. In part, this is because the BB configurations are designed with careful forethought regarding the intended orbit so that BB marker projections do not overlap in projection views. Examples include helical arrangements of BBs (Rougee et al 1993 Proc. SPIE 1897 161-9) such that markers do not overlap in projections acquired from a circular orbit and circular arrangements of BBs (Cho et al 2005 Med. Phys. 32 968-83). As a more general alternative, this work proposes a calibration method based on an array of line-shaped, radio-opaque wire segments. With this method, geometric parameter estimation is accomplished by relating the 3D line equations representing the wires to the 2D line equations of their projections. The use of line fiducials simplifies many challenges with fiducial recognition and extraction in an orbit-independent manner. For example, their projections can overlap only mildly, for any gantry pose, as long as the wires are mutually non-coplanar in 3D. The method was tested in application to circular and non-circular trajectories in simulation and in real orbits executed using a mobile C-arm prototype for cone-beam CT. Results indicated high calibration accuracy, as measured by forward and backprojection/triangulation error metrics. Triangulation errors on the order of microns and backprojected ray deviations uniformly less than 0.2 mm were observed in both real and simulated orbits. Mean forward projection errors less than 0.1 mm were observed in a comprehensive sweep of different C-arm gantry angulations. Finally, successful integration of the method into a CT imaging chain was demonstrated in head phantom scans.",
keywords = "BB phantom, C-arm, cone-beam CT, geometric calibration, line fiducial, wire phantom",
author = "Jacobson, {M. W.} and Ketcha, {M. D.} and S. Capostagno and A. Martin and A. Uneri and J. Goerres and {De Silva}, T. and S. Reaungamornrat and R. Han and A. Manbachi and Stayman, {J. W.} and S. Vogt and G. Kleinszig and Siewerdsen, {J. H.}",
year = "2018",
month = "1",
day = "1",
doi = "10.1088/1361-6560/aa9910",
language = "English (US)",
volume = "63",
journal = "Physics in Medicine and Biology",
issn = "0031-9155",
publisher = "IOP Publishing Ltd.",
number = "2",

}

TY - JOUR

T1 - A line fiducial method for geometric calibration of cone-beam CT systems with diverse scan trajectories

AU - Jacobson, M. W.

AU - Ketcha, M. D.

AU - Capostagno, S.

AU - Martin, A.

AU - Uneri, A.

AU - Goerres, J.

AU - De Silva, T.

AU - Reaungamornrat, S.

AU - Han, R.

AU - Manbachi, A.

AU - Stayman, J. W.

AU - Vogt, S.

AU - Kleinszig, G.

AU - Siewerdsen, J. H.

PY - 2018/1/1

Y1 - 2018/1/1

N2 - Modern cone-beam CT systems, especially C-arms, are capable of diverse source-detector orbits. However, geometric calibration of these systems using conventional configurations of spherical fiducials (BBs) may be challenged for novel source-detector orbits and system geometries. In part, this is because the BB configurations are designed with careful forethought regarding the intended orbit so that BB marker projections do not overlap in projection views. Examples include helical arrangements of BBs (Rougee et al 1993 Proc. SPIE 1897 161-9) such that markers do not overlap in projections acquired from a circular orbit and circular arrangements of BBs (Cho et al 2005 Med. Phys. 32 968-83). As a more general alternative, this work proposes a calibration method based on an array of line-shaped, radio-opaque wire segments. With this method, geometric parameter estimation is accomplished by relating the 3D line equations representing the wires to the 2D line equations of their projections. The use of line fiducials simplifies many challenges with fiducial recognition and extraction in an orbit-independent manner. For example, their projections can overlap only mildly, for any gantry pose, as long as the wires are mutually non-coplanar in 3D. The method was tested in application to circular and non-circular trajectories in simulation and in real orbits executed using a mobile C-arm prototype for cone-beam CT. Results indicated high calibration accuracy, as measured by forward and backprojection/triangulation error metrics. Triangulation errors on the order of microns and backprojected ray deviations uniformly less than 0.2 mm were observed in both real and simulated orbits. Mean forward projection errors less than 0.1 mm were observed in a comprehensive sweep of different C-arm gantry angulations. Finally, successful integration of the method into a CT imaging chain was demonstrated in head phantom scans.

AB - Modern cone-beam CT systems, especially C-arms, are capable of diverse source-detector orbits. However, geometric calibration of these systems using conventional configurations of spherical fiducials (BBs) may be challenged for novel source-detector orbits and system geometries. In part, this is because the BB configurations are designed with careful forethought regarding the intended orbit so that BB marker projections do not overlap in projection views. Examples include helical arrangements of BBs (Rougee et al 1993 Proc. SPIE 1897 161-9) such that markers do not overlap in projections acquired from a circular orbit and circular arrangements of BBs (Cho et al 2005 Med. Phys. 32 968-83). As a more general alternative, this work proposes a calibration method based on an array of line-shaped, radio-opaque wire segments. With this method, geometric parameter estimation is accomplished by relating the 3D line equations representing the wires to the 2D line equations of their projections. The use of line fiducials simplifies many challenges with fiducial recognition and extraction in an orbit-independent manner. For example, their projections can overlap only mildly, for any gantry pose, as long as the wires are mutually non-coplanar in 3D. The method was tested in application to circular and non-circular trajectories in simulation and in real orbits executed using a mobile C-arm prototype for cone-beam CT. Results indicated high calibration accuracy, as measured by forward and backprojection/triangulation error metrics. Triangulation errors on the order of microns and backprojected ray deviations uniformly less than 0.2 mm were observed in both real and simulated orbits. Mean forward projection errors less than 0.1 mm were observed in a comprehensive sweep of different C-arm gantry angulations. Finally, successful integration of the method into a CT imaging chain was demonstrated in head phantom scans.

KW - BB phantom

KW - C-arm

KW - cone-beam CT

KW - geometric calibration

KW - line fiducial

KW - wire phantom

UR - http://www.scopus.com/inward/record.url?scp=85040951619&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85040951619&partnerID=8YFLogxK

U2 - 10.1088/1361-6560/aa9910

DO - 10.1088/1361-6560/aa9910

M3 - Article

C2 - 29116058

AN - SCOPUS:85040951619

VL - 63

JO - Physics in Medicine and Biology

JF - Physics in Medicine and Biology

SN - 0031-9155

IS - 2

M1 - 025030

ER -