A finite element study of the initiation of failure of fixation in cemented femoral total hip components

Timothy Harrigan, J. A. Kareh, D. O. O'Connor, D. W. Burke, W. H. Harris

Research output: Contribution to journalArticle

Abstract

In order to study initial mechanisms of failure in cemented femoral total hip components, an anatomically accurate three‐dimensional linear finite element model was constructed and verified against experimental strain measurements in the cement mantle. Good agreement was found between predicted and measured strains. The likelihood of failure initiation due to cement‐prosthesis debonding and crack initiation at voids was studied for loading conditions simulating both one‐legged stance and stair climbing. The “out of plane” forces involved in stair climbing appear to be the greatest threat to the fixation of total hip replacements. In stair climbing, cement‐prosthesis debonding and pore crack initiation were probable in the proximal anteromedial region of the cement mantle, and near the distal tip of the implant. The proximal stresses in stair climbing were higher than the distal stresses in either stair climbing or one‐legged stance.

Original languageEnglish (US)
Pages (from-to)134-144
Number of pages11
JournalJournal of Orthopaedic Research
Volume10
Issue number1
DOIs
StatePublished - Jan 1 1992
Externally publishedYes

Fingerprint

Thigh
Hip
Hip Replacement Arthroplasties
Stair Climbing

Keywords

  • Cement
  • Debonding
  • Failure
  • Interface
  • Stair climbing
  • Total hip

ASJC Scopus subject areas

  • Orthopedics and Sports Medicine

Cite this

A finite element study of the initiation of failure of fixation in cemented femoral total hip components. / Harrigan, Timothy; Kareh, J. A.; O'Connor, D. O.; Burke, D. W.; Harris, W. H.

In: Journal of Orthopaedic Research, Vol. 10, No. 1, 01.01.1992, p. 134-144.

Research output: Contribution to journalArticle

Harrigan, Timothy ; Kareh, J. A. ; O'Connor, D. O. ; Burke, D. W. ; Harris, W. H. / A finite element study of the initiation of failure of fixation in cemented femoral total hip components. In: Journal of Orthopaedic Research. 1992 ; Vol. 10, No. 1. pp. 134-144.
@article{a257d7c3ed3349649ec2fdcdabec2195,
title = "A finite element study of the initiation of failure of fixation in cemented femoral total hip components",
abstract = "In order to study initial mechanisms of failure in cemented femoral total hip components, an anatomically accurate three‐dimensional linear finite element model was constructed and verified against experimental strain measurements in the cement mantle. Good agreement was found between predicted and measured strains. The likelihood of failure initiation due to cement‐prosthesis debonding and crack initiation at voids was studied for loading conditions simulating both one‐legged stance and stair climbing. The “out of plane” forces involved in stair climbing appear to be the greatest threat to the fixation of total hip replacements. In stair climbing, cement‐prosthesis debonding and pore crack initiation were probable in the proximal anteromedial region of the cement mantle, and near the distal tip of the implant. The proximal stresses in stair climbing were higher than the distal stresses in either stair climbing or one‐legged stance.",
keywords = "Cement, Debonding, Failure, Interface, Stair climbing, Total hip",
author = "Timothy Harrigan and Kareh, {J. A.} and O'Connor, {D. O.} and Burke, {D. W.} and Harris, {W. H.}",
year = "1992",
month = "1",
day = "1",
doi = "10.1002/jor.1100100116",
language = "English (US)",
volume = "10",
pages = "134--144",
journal = "Journal of Orthopaedic Research",
issn = "0736-0266",
publisher = "John Wiley and Sons Inc.",
number = "1",

}

TY - JOUR

T1 - A finite element study of the initiation of failure of fixation in cemented femoral total hip components

AU - Harrigan, Timothy

AU - Kareh, J. A.

AU - O'Connor, D. O.

AU - Burke, D. W.

AU - Harris, W. H.

PY - 1992/1/1

Y1 - 1992/1/1

N2 - In order to study initial mechanisms of failure in cemented femoral total hip components, an anatomically accurate three‐dimensional linear finite element model was constructed and verified against experimental strain measurements in the cement mantle. Good agreement was found between predicted and measured strains. The likelihood of failure initiation due to cement‐prosthesis debonding and crack initiation at voids was studied for loading conditions simulating both one‐legged stance and stair climbing. The “out of plane” forces involved in stair climbing appear to be the greatest threat to the fixation of total hip replacements. In stair climbing, cement‐prosthesis debonding and pore crack initiation were probable in the proximal anteromedial region of the cement mantle, and near the distal tip of the implant. The proximal stresses in stair climbing were higher than the distal stresses in either stair climbing or one‐legged stance.

AB - In order to study initial mechanisms of failure in cemented femoral total hip components, an anatomically accurate three‐dimensional linear finite element model was constructed and verified against experimental strain measurements in the cement mantle. Good agreement was found between predicted and measured strains. The likelihood of failure initiation due to cement‐prosthesis debonding and crack initiation at voids was studied for loading conditions simulating both one‐legged stance and stair climbing. The “out of plane” forces involved in stair climbing appear to be the greatest threat to the fixation of total hip replacements. In stair climbing, cement‐prosthesis debonding and pore crack initiation were probable in the proximal anteromedial region of the cement mantle, and near the distal tip of the implant. The proximal stresses in stair climbing were higher than the distal stresses in either stair climbing or one‐legged stance.

KW - Cement

KW - Debonding

KW - Failure

KW - Interface

KW - Stair climbing

KW - Total hip

UR - http://www.scopus.com/inward/record.url?scp=0026580011&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0026580011&partnerID=8YFLogxK

U2 - 10.1002/jor.1100100116

DO - 10.1002/jor.1100100116

M3 - Article

C2 - 1727933

AN - SCOPUS:0026580011

VL - 10

SP - 134

EP - 144

JO - Journal of Orthopaedic Research

JF - Journal of Orthopaedic Research

SN - 0736-0266

IS - 1

ER -