A dynamic random dot stereogram-based system for strabismus and amblyopia screening of infants and young children

Kurt A. Simons, Allen Moss

Research output: Contribution to journalArticle

Abstract

A hardware and software system is described that administers a dynamic form of random dot stereogram (RDS) vision test to subjects of any age for purposes of detecting the presence of strabismus (e.g. 'crossed eyes') or amblyopia ('lazy eye'). The dynamic RDS is generated by an image processor under host computer control and is viewed on a color television monitor. Stereoscopic viewing is achieved through presentation of one image of the stereo pair in the red channel and the other in the green channel of the monitor, and having the subject view the monitor through a red and green window. Response is in the form of eye movement, detected by means of a television-based eye tracker, the output of which is fed to the host computer for analysis.

Original languageEnglish (US)
Pages (from-to)33-46
Number of pages14
JournalComputers in Biology and Medicine
Volume11
Issue number1
DOIs
StatePublished - 1981
Externally publishedYes

Fingerprint

Amblyopia
Strabismus
Television
Screening
Vision Tests
Color television
Eye movements
Computer control
Eye Movements
Software
Color
Hardware

Keywords

  • Amblyopia
  • Child
  • Eye movements
  • Image processor
  • Infant
  • Random dot stereogram
  • Strabismus
  • Television
  • Vision testing

ASJC Scopus subject areas

  • Computer Science Applications

Cite this

@article{c954f7da5e2048f8b71889a929b43691,
title = "A dynamic random dot stereogram-based system for strabismus and amblyopia screening of infants and young children",
abstract = "A hardware and software system is described that administers a dynamic form of random dot stereogram (RDS) vision test to subjects of any age for purposes of detecting the presence of strabismus (e.g. 'crossed eyes') or amblyopia ('lazy eye'). The dynamic RDS is generated by an image processor under host computer control and is viewed on a color television monitor. Stereoscopic viewing is achieved through presentation of one image of the stereo pair in the red channel and the other in the green channel of the monitor, and having the subject view the monitor through a red and green window. Response is in the form of eye movement, detected by means of a television-based eye tracker, the output of which is fed to the host computer for analysis.",
keywords = "Amblyopia, Child, Eye movements, Image processor, Infant, Random dot stereogram, Strabismus, Television, Vision testing",
author = "Simons, {Kurt A.} and Allen Moss",
year = "1981",
doi = "10.1016/0010-4825(81)90014-7",
language = "English (US)",
volume = "11",
pages = "33--46",
journal = "Computers in Biology and Medicine",
issn = "0010-4825",
publisher = "Elsevier Limited",
number = "1",

}

TY - JOUR

T1 - A dynamic random dot stereogram-based system for strabismus and amblyopia screening of infants and young children

AU - Simons, Kurt A.

AU - Moss, Allen

PY - 1981

Y1 - 1981

N2 - A hardware and software system is described that administers a dynamic form of random dot stereogram (RDS) vision test to subjects of any age for purposes of detecting the presence of strabismus (e.g. 'crossed eyes') or amblyopia ('lazy eye'). The dynamic RDS is generated by an image processor under host computer control and is viewed on a color television monitor. Stereoscopic viewing is achieved through presentation of one image of the stereo pair in the red channel and the other in the green channel of the monitor, and having the subject view the monitor through a red and green window. Response is in the form of eye movement, detected by means of a television-based eye tracker, the output of which is fed to the host computer for analysis.

AB - A hardware and software system is described that administers a dynamic form of random dot stereogram (RDS) vision test to subjects of any age for purposes of detecting the presence of strabismus (e.g. 'crossed eyes') or amblyopia ('lazy eye'). The dynamic RDS is generated by an image processor under host computer control and is viewed on a color television monitor. Stereoscopic viewing is achieved through presentation of one image of the stereo pair in the red channel and the other in the green channel of the monitor, and having the subject view the monitor through a red and green window. Response is in the form of eye movement, detected by means of a television-based eye tracker, the output of which is fed to the host computer for analysis.

KW - Amblyopia

KW - Child

KW - Eye movements

KW - Image processor

KW - Infant

KW - Random dot stereogram

KW - Strabismus

KW - Television

KW - Vision testing

UR - http://www.scopus.com/inward/record.url?scp=0019350845&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0019350845&partnerID=8YFLogxK

U2 - 10.1016/0010-4825(81)90014-7

DO - 10.1016/0010-4825(81)90014-7

M3 - Article

C2 - 7014078

AN - SCOPUS:0019350845

VL - 11

SP - 33

EP - 46

JO - Computers in Biology and Medicine

JF - Computers in Biology and Medicine

SN - 0010-4825

IS - 1